Ion detection of 2, 4-dinitroanisole by ion chromatography

Si Yu Wu, Qiao Zhu, Min Xue, Zhi Qun Chen, Min Xu, Lin Jia, Min Chang Wang, Zhi Hui Lin, Zi Hui Meng*

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

摘要

To determine the trace ions in energetic compound 2, 4-dinitroanisole (DNAN) sample, the qualitative and quantitative analysis of ions in DNAN sample were performed by an ion chromatography method. Results show that the sample contains 8 kinds of ions of Li+, Na+, NH4+, Mg2+, Ca2+, Cl-, NO3-, SO42-. Their mass fractions are 13.2348, 70.4080, 15.3270, 83.2659, 154.1580, 156.4740, 21.6544, 210.3100μg/mL, respectively. The method has the advantages of rapid detection and reliable quantification.

源语言英语
页(从-至)73-76
页数4
期刊Huozhayao Xuebao/Chinese Journal of Explosives and Propellants
38
3
DOI
出版状态已出版 - 1 6月 2015

指纹

探究 'Ion detection of 2, 4-dinitroanisole by ion chromatography' 的科研主题。它们共同构成独一无二的指纹。

引用此

Wu, S. Y., Zhu, Q., Xue, M., Chen, Z. Q., Xu, M., Jia, L., Wang, M. C., Lin, Z. H., & Meng, Z. H. (2015). Ion detection of 2, 4-dinitroanisole by ion chromatography. Huozhayao Xuebao/Chinese Journal of Explosives and Propellants, 38(3), 73-76. https://doi.org/10.14077/j.issn.1007-7812.2015.03.014