Microcomputer-controlled scanning Fabry-Perot interferometer

Weirui Zhao*, Naiwen He, Genrui Cao

*此作品的通讯作者

科研成果: 期刊稿件会议文章同行评审

3 引用 (Scopus)

摘要

Using the technique of capacitance micrometry it is possible to measure very small displacements. Here, a microcomputer-controlled scanning Fabry-Perot interferometer is presented. In each scanning step, the parallelism of Fabry-Perot interferometer is monitored and adjusted in real-time with the aid of three capacitance micrometers - it is the feature of this system, where the readjustment and scanning are realized with three electrostrictive actuators, and the parallelism and the spacing of the mirrors are controlled by three capacitance micrometers. The light information from the interferometer is detected by a photo-detector. All of these are controlled by a microcomputer in order to realize the scanning and real-time control of Fabry-Perot interferometer. This system removes the problem of the nonlinear response and hysteresis associated with electrostrictive actuators and allows precise control of both parallelism and spacing of the mirrors.

源语言英语
页(从-至)515-522
页数8
期刊Proceedings of SPIE - The International Society for Optical Engineering
3786
出版状态已出版 - 1999
活动Proceedings of the 1999 Optomechanical Engineering and Vibration Control - Denver, CO, USA
期限: 20 7月 199923 7月 1999

指纹

探究 'Microcomputer-controlled scanning Fabry-Perot interferometer' 的科研主题。它们共同构成独一无二的指纹。

引用此