@inproceedings{d938788fdea442ea90f1d858faa3c620,
title = "Spectral calibration method for Stokes polarimeter",
abstract = "An efficient method that allows spectral calibration for a Stokes polarimeter is presented in this paper. The calibration procedure only uses a reference polarizer to generate an arbitrary linear polarization state. The reference polarizer could also be calibrated while calibrating the Stokes polarimeter. In addition, this method does not involve the small-angle approximation and could avoid the influence of the initial azimuth error of the retarder and the reference polarizer. The experiment was carried out by a spectroscopy and a polarization state analyzer with a rotating wave plate. Experimental results show that the alignments of axis as well as the true retardation at a specific wavelength of the retarder are obtained.",
keywords = "Calibration, Polarimeter, Stokes",
author = "Guodong Zhou and Jianhui Li and Yanqiu Li and Yuanhe Li",
note = "Publisher Copyright: {\textcopyright} 2021 COPYRIGHT SPIE.; 12th International Conference on Information Optics and Photonics, CIOP 2021 ; Conference date: 23-07-2021 Through 26-07-2021",
year = "2021",
doi = "10.1117/12.2603827",
language = "English",
series = "Proceedings of SPIE - The International Society for Optical Engineering",
publisher = "SPIE",
editor = "Yue Yang",
booktitle = "Twelfth International Conference on Information Optics and Photonics, CIOP 2021",
address = "United States",
}