Research on calibration method of three band infrared integrated radiometer

Xia Wang, Zhi Yun Gao, Jianyong Zhang, Weiqi Jin

Research output: Contribution to journalConference articlepeer-review

4 Citations (Scopus)

Abstract

A three band infrared integrated radiometer was developed for field measurement. The composition and work theory of the radiometer are described in this paper. The detailed calibration methods are analyzed for extended source and point source, and corresponding measurement results are given. With test data, the experimental results and instrument performance are analyzed in detail, in respect of the equivalent temperature of inter reference blackbody, linearity with target temperature, measurement error of point source and extended source. To lessen the influents of calibration for measurement, the attentive problems in actual calibration of extended source and point source are summarized, and satisfied results are acquired.

Original languageEnglish
Pages (from-to)133-138
Number of pages6
JournalProceedings of SPIE - The International Society for Optical Engineering
Volume4927
DOIs
Publication statusPublished - 20 Sept 2002
EventOptical Design and Testing 2002 - Shanghai, China
Duration: 14 Oct 200218 Oct 2002

Keywords

  • Calibration
  • Infrared
  • Integrated radiometer

Fingerprint