Abstract
Minimally invasive vascular interventional surgery is widely used and remote-controlled vascular interventional surgery robots (RVIRs) are being developed to reduce the occupational risk of the intervening physician in minimally invasive vascular interventional surgeries. Skilled surgeon performs surgeries mainly depending on the detection of collisions. Inaccurate force feedback will be difficult for surgeons to perform surgeries or even results in medical accidents. In addition, the surgeon cannot quickly and easily distinguish whether the proximal force exceeds the safety threshold of blood vessels or not, and thus it results in damage to the blood vessels. In this paper, we present a novel method comprising compensatory force measurement and multimodal force feedback (MFF). Calibration experiments and performance evaluation experiments were carried out. Experimental results demonstrated that the proposed method can measure the proximal force of catheter/guidewire accurately and assist surgeons to distinguish the change of proximal force more easily. This novel method is suitable for use in actual surgical operations.
Original language | English |
---|---|
Article number | 74 |
Journal | Biomedical Microdevices |
Volume | 20 |
Issue number | 3 |
DOIs | |
Publication status | Published - 1 Sept 2018 |
Keywords
- Force feedback
- Force measurement
- Minimally invasive vascular interventional surgery
- Remote-controlled vascular interventional robot (RVIR)