Error analysis of the calibration of a dual-rotating-retarder Mueller matrix polarimeter

Xuemin Cheng*, Maolin Li, Jialing Zhou, Hui Ma, Qun Hao

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

17 引用 (Scopus)

摘要

We previously developed a Mueller matrix microscope by combining a dual-rotating-retarder device with a commercial microscope, which holds promise for pathologic diagnosis applications. In practical applications, the signals to be measured are sensitive to systematic errors, and the measurement accuracy and the instrument calibration affect the capacity to characterize the data. In conventional calibration, the initial positions of the rotating devices are constrained, causing calibration instability. To improve the calibration, we examine the mathematical model of the calibration process using previous studies, and derive a calibration algorithm with reduced error magnification. Our experimental results show the proposed algorithm improves calibration accuracy, and provides more robust results.

源语言英语
页(从-至)7067-7074
页数8
期刊Applied Optics
56
25
DOI
出版状态已出版 - 1 9月 2017

指纹

探究 'Error analysis of the calibration of a dual-rotating-retarder Mueller matrix polarimeter' 的科研主题。它们共同构成独一无二的指纹。

引用此