Endoscopic swept-source optical coherence tomography based on a two-axis microelectromechanical system mirror

Donglin Wang, Linlai Fu, Xin Wang, Zhongjian Gong*, Sean Samuelson, Can Duan, Hongzhi Jia, Jun Shan Ma, Huikai Xie

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

44 引用 (Scopus)

摘要

A microelectromechanical system (MEMS) mirror based endoscopic swept-source optical coherence tomography (SS-OCT) system that can perform three-dimensional (3-D) imaging at high speed is reported. The key component enabling 3-D endoscopic imaging is a two-axis MEMS scanning mirror which has a 0.8 × 0.8 mm2mirror plate and a 1.6 × 1.4 mm 2 device footprint. The diameter of the endoscopic probe is only 3.5 mm. The imaging rate of the SS-OCT system is 50 frames/s. OCT images of both human suspicious oral leukoplakia tissue and normal buccal mucosa were taken in vivo and compared. The OCT imaging result agrees well with the histopathological analysis.