Design of a Multi-Parameter Fusion Sensor and System for Respiratory Monitoring of Mechanically Ventilated Patients in the ICU

Shuai Ren*, Xiaohan Wang, Maolin Cai, Yan Shi, Tao Wang, Zujin Luo*

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

摘要

In order to achieve precise respiratory therapy for mechanically ventilated patients, real-time monitoring of the state parameters of inhaled and exhaled gases is required. These parameters are primarily measured by ventilators, with limitations such as insufficient monitoring parameters, circuit leaks, and constraints imposed by distance and obstacles. This paper designs a low-power wireless sensor for multi-parameter monitoring near the patient, which can be used continuously for approximately 60 days. Based on this sensor, an intelligent respiratory monitoring system with a distributed architecture is proposed to achieve intelligent patient-ventilator asynchrony (PVA) perception. Experimental results show that the system can stably and accurately collect and transmit data, with measurement errors for pressure, flow, temperature, humidity, and CO2 concentration being pm1.3%, pm2.1%, pm 0.6° C, pm1% RH, pm0.3 mmHg respectively. The proposed sensor and system have the potential to enhance the efficiency and intelligence of medical care significantly.

源语言英语
期刊IEEE Journal of Biomedical and Health Informatics
DOI
出版状态已接受/待刊 - 2024

指纹

探究 'Design of a Multi-Parameter Fusion Sensor and System for Respiratory Monitoring of Mechanically Ventilated Patients in the ICU' 的科研主题。它们共同构成独一无二的指纹。

引用此