摘要
This paper reports a single-crystalline silicon (SCS) micromirror used for laser beam scanning in an endoscopic optical coherence tomography (OCT) system. The micromirror is fabricated by using a deep reactive-ion-etch (DRIE) post-CMOS micromachining process. Thin bimorph actuation structures and movable bulk silicon structures are simultaneously achieved. The micromirror is 1 mm by 1 mm in size, coated with aluminum, and thermally actuated by an integrated polysilicon heater. The radius of curvature of the mirror surface is 50 cm. The mirror rotates 17° when a 15 mA current is applied. Cross-sectional images of 500×1000 pixels covering an area of 2.9 mm by 2.8 mm are acquired at 5 frames/s by using an OCT system based on this micromirror.
源语言 | 英语 |
---|---|
页 | 495-498 |
页数 | 4 |
出版状态 | 已出版 - 2002 |
已对外发布 | 是 |
活动 | 15th IEEE International Conference on Micro Electro Mechanical Systems MEMS 2002 - Las Vegas, NV, 美国 期限: 20 1月 2002 → 24 1月 2002 |
会议
会议 | 15th IEEE International Conference on Micro Electro Mechanical Systems MEMS 2002 |
---|---|
国家/地区 | 美国 |
市 | Las Vegas, NV |
时期 | 20/01/02 → 24/01/02 |