A Method to Model the X-ray Response of Electronic Portal Imaging Device

Jun Zhang, Ziting Fan, Ning Li, Junhai Wen*

*此作品的通讯作者

科研成果: 书/报告/会议事项章节会议稿件同行评审

摘要

Electronic Portal Imaging Device (EPID) has been widely used in dose verification of radiotherapy. When using EPID for dose verification, it is necessary to calculate the primary and scatter rays in the EPID plane respectively. In our study, we use a series of measurement data to model the scatter to primary ratio (SPR) of EPID under different conditions. It is found that the SPR of EPID is related to the field size, phantom thickness, exit gap, and off-axis distance. In the process of treatment, the primary and scatter values of each point in EPID can be quickly calculated by the corresponding SPR. This method simplifies the modeling process of primary and scatter rays of EPID and improves the calculation accuracy of off-axis points.

源语言英语
主期刊名2021 6th International Conference on Signal and Image Processing, ICSIP 2021
出版商Institute of Electrical and Electronics Engineers Inc.
411-415
页数5
ISBN(电子版)9780738133737
DOI
出版状态已出版 - 2021