摘要
In this paper, a high-sensitivity orthogonal receiving system was designed based on finite-unit analysis to improve MPI signal detection sensitivity and magnetic field homogeneity for preclinical MPI of small animal imaging. The orthogonal receiving coil set was designed to add a new x-direction receiving coil, to improve the reception sensitivity and optimize imaging field of view area in z-direction. Simulation experiments, construction and calibration were carried out in the preclinical MPI platform with the high sensitivity orthogonal receiving systems to perform phantom imaging and small animal imaging experiments. Results show that the new orthogonal system can improve the image quality and contrast-noise rate effectively.
投稿的翻译标题 | High Sensitivity Orthogonal Receiving System for Preclinical Magnetic Particle Imaging |
---|---|
源语言 | 繁体中文 |
页(从-至) | 312-317 |
页数 | 6 |
期刊 | Beijing Ligong Daxue Xuebao/Transaction of Beijing Institute of Technology |
卷 | 42 |
期 | 3 |
DOI | |
出版状态 | 已出版 - 1 3月 2022 |
关键词
- Homogeneity
- Magnetic particle imaging
- Orthogonal coils
- Sensitivity