摘要
A wide-area fundus camera used for screening the retinae of infants was designed. In this paper, the design methods of the device's illuminating and imaging systems were investigated. Based on James Polans' wide-field human eye model and a set of ophthalmic anatomy data, an infant eye model was established. Then, a tapered fiber scheme was proposed for wide area illumination. Finally, the design method of a wide-area fundus camera imaging system, including the contact lens and relay lens, is introduced. The design example shows that the Field Of View (FOV) of the wide-area fundus camera can reach 130°, and the object resolution of the fundus can reach 10 μm. The design results meet the national standards YY0634-2008 for fundus imaging equipment and meet the requirements for infant retina screening.
投稿的翻译标题 | Optical system design and simulation of a wide-area fundus camera |
---|---|
源语言 | 繁体中文 |
页(从-至) | 814-821 |
页数 | 8 |
期刊 | Chinese Optics |
卷 | 13 |
期 | 4 |
DOI | |
出版状态 | 已出版 - 1 8月 2020 |
关键词
- Medical optics
- Optical design
- Wide-area retinal imaging