Solid phase microextraction device coupled with miniature mass spectrometry and mathematical model of its ion chronogram

Hao Gu*, Jiwen Li*, Qiong Liang, Wei Xu

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

3 引用 (Scopus)

摘要

Modern solid phase microextraction (SPME) device linked with mass spectrometry (SPME-MS) has evolved from producing ion chronogram as flat noisy signal to as unimodal-like signal. We designed a SPME device, which is closer in morphology to LC column, linked it with a miniature mass spectrometer (SPME-Mini MS), and proposed a mathematical model that elution of compound from the SPME device is equivalent to overlay of elution of the compound from the infinite LC columns with the lengths between 0 and the length of the device and it can generate an ion chronogram as right-skew unimodal signal. Rhodamine B as analyte was used for experimental verification and its unimodal signal was used to fit the parameters of a computer simulation program based on the model. The experimental results and simulations empirically cross-confirmed that SPME-Mini MS can generate ion chronogram as clean right-skew unimodal signal. Furthermore, the SPME-Mini MS system was used for quantitative analysis of psychotropic drugs (i.e. risperidone and aripiprazole) in artificial urine. The results preliminarily demonstrated that the system can utilize area under unimodal signal for quantitative analysis and has potential to be applied for on-site, fast and accurate quantification of drugs and other compounds.