Research on accurate non-contact temperature measurement method for telescope mirror

Shanjie Huang, Lingxue Wang, Xu Hu, Shouzhang Yuan, Fangyu Xu, Jinsong Zhao, Yi Cai

科研成果: 期刊稿件文章同行评审

1 引用 (Scopus)

摘要

Non-contact temperature measurement for a solar telescope mirror is critical for improving the mirror seeing and thermal deformation of solar telescopes, a long-standing challenge in astronomy. This challenge arises from the telescope mirror’s inherent weak thermal radiation, often overwhelmed by reflected background radiations due to its high reflectivity. In this work, an infrared mirror thermometer (IMT) is equipped with a thermally-modulated reflector, and a measurement method based on an equation for extracting mirror radiation (EEMR) has been developed for probing the accurate radiation and temperature of the telescope mirror. Using this approach, we can extract the mirror radiation from the instrumental background radiation via the EEMR. This reflector has been designed to amplify the mirror radiation signal incident on the infrared sensor of IMT, while inhibiting the radiation noise from the ambient environment. In addition, we also propose a set of evaluation methods for IMT performance based on EEMR. The results reveal that the temperature measurement accuracy of IMT to the solar telescope mirror using this measurement method can be achieved better than ±0.15°C.

源语言英语
页(从-至)21521-21541
页数21
期刊Optics Express
31
13
DOI
出版状态已出版 - 19 6月 2023

指纹

探究 'Research on accurate non-contact temperature measurement method for telescope mirror' 的科研主题。它们共同构成独一无二的指纹。

引用此