摘要
We propose an inter-Ascan speckle decorrelation based method that can quantitatively assess blood flow normal to the direction of the optical coherence tomography imaging beam. To validate this method, we performed a systematic study using both phantom and in vivo animal models. Results show that our speckle analysis method can accurately extract transverse flow speed with high spatial and temporal resolution.
源语言 | 英语 |
---|---|
页(从-至) | 805-807 |
页数 | 3 |
期刊 | Optics Letters |
卷 | 38 |
期 | 5 |
DOI | |
出版状态 | 已出版 - 1 3月 2013 |
已对外发布 | 是 |