Portable and multiplexed lateral flow immunoassay reader based on SERS for highly sensitive point-of-care testing

Rui Xiao, Luchun Lu, Zhen Rong, Chongwen Wang*, Yongjin Peng, Feng Wang, Jinhai Wang, Meijie Sun, Jian Dong, Dongfeng Wang, Luanluan Wang, Nanxi Sun, Shengqi Wang*

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

90 引用 (Scopus)

摘要

A portable surface-enhanced Raman scattering (SERS)-based lateral flow immunoassay (LFIA) reader with multiplexed detection was developed using an integrated LFIA reaction column. The proposed LFIA reader was designed to simultaneously detect multiple samples or samples with multiple biomarkers. With the integrated LFIA reaction column, we achieved the specific detection of alpha-fetoprotein (AFP), carcinoembryonic antigen (CEA), and prostate-specific antigen (PSA) with a detection limit of 0.01 ng/mL, which was three orders of magnitude lower than that of the visual signal. We also investigated the uniformity of channels based on an eight-channel integrated LFIA reaction column. The relative standard deviation values of the SERS intensity of the eight-channel for measuring the AFP, CEA, and PSA antigens at 1323 cm−1 were 13%, 4.8%, and 5%, respectively. We detected 45 clinical serum samples of the three antigens using the proposed portable SERS-based LFIA reader to further confirm its applicability to clinical samples. The SERS signals of the positive sera were higher than those of the negative sera and their thrice standard deviation. This result indicated the practicality of the developed integrated reaction column and the proposed portable and multiplexed Raman reader. This work provides a new high-sensitivity, multiplexed, and automated SERS-based LFIA detector for use in the point-of-care setting.

源语言英语
文章编号112524
期刊Biosensors and Bioelectronics
168
DOI
出版状态已出版 - 15 11月 2020
已对外发布

指纹

探究 'Portable and multiplexed lateral flow immunoassay reader based on SERS for highly sensitive point-of-care testing' 的科研主题。它们共同构成独一无二的指纹。

引用此