Novel ultra-thin near-eye augmented-view device (UNA-)

Y. Zhu, J. Chang, Y. J. Niu, W. L. Chen, X. Y. Du

科研成果: 期刊稿件文章同行评审

2 引用 (Scopus)

摘要

The current augmented-view device for glaucoma patients are limited by the low angle minification (less than 3X), image overlapping and pupil mismatching. We present a novel ultra-thin near-eye augmented-view device (UNAD) with a proper minification angle of 4X to avoid scanning eye movements. The device is realized by one symmetrical ultra-thin offaxis eight-mirror reversed telescopic optical system (field-of-view (FOV): x:-34°∼3°; y: -5°∼-20°, 5°∼20°) and one central un-minified optical system (FOV: -5°∼5°), and hence, the system can achieve a FOV (x:-34°∼3°; y: -20°∼20°) without overlapping images. Furthermore, the device fully accounts for pupil matching with the entrance pupil of the human eye. Finally, the simulation results in CODE V and LightTools verify that our proposed system is feasible and has potential practical value for glaucoma patients.

源语言英语
页(从-至)1305-1313
页数9
期刊Optics Express
24
2
DOI
出版状态已出版 - 25 1月 2016

指纹

探究 'Novel ultra-thin near-eye augmented-view device (UNA-)' 的科研主题。它们共同构成独一无二的指纹。

引用此