Method of Oxygen Saturation Detection for Wearable Smart Mobile Device

Bo Wen Li, Qun Wang*, Zhi Wen Liu

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

摘要

A fast method of blood oxygen saturation detection for wearable smart mobile devices was proposed. Based on the no-correlated relationship between the pulse component of photoplethysmography and the noise component, the noise can be suppressed and the blood oxygen saturation can be calculated. The method has been verified by processing the low perfusion signal from the blood oxygen simulator. The results show that this method has a small amount of calculation with good anti-noise ability. It can accurately detect the blood oxygen saturation from weak signals with noise.

源语言英语
页(从-至)136-140
页数5
期刊Beijing Ligong Daxue Xuebao/Transaction of Beijing Institute of Technology
37
出版状态已出版 - 1 6月 2017

指纹