@inproceedings{04903a7081cc406db9e5fa23cb0ecf0c,
title = "Injection-depth-locking axial motion guided handheld micro-injector using CP-SSOCT",
abstract = "This paper presents a handheld micro-injector system using common-path swept source optical coherence tomography (CP-SSOCT) as a distal sensor with highly accurate injection-depth-locking. To achieve real-time, highly precise, and intuitive freehand control, the system used graphics processing unit (GPU) to process the oversampled OCT signal with high throughput and a smart customized motion monitoring control algorithm. A performance evaluation was conducted with 60-insertions and fluorescein dye injection tests to show how accurately the system can guide the needle and lock to the target depth. The evaluation tests show our system can guide the injection needle into the desired depth with 4.12um average deviation error while injecting 50nl of fluorescein dye.",
author = "Cheon, {Gyeong Woo} and Yong Huang and Kwag, {Hye Rin} and Kim, {Ki Young} and Taylor, {Russell H.} and Gehlbach, {Peter L.} and Kang, {Jin U.}",
note = "Publisher Copyright: {\textcopyright} 2014 IEEE.; 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2014 ; Conference date: 26-08-2014 Through 30-08-2014",
year = "2014",
month = nov,
day = "2",
doi = "10.1109/EMBC.2014.6945036",
language = "English",
series = "2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2014",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
pages = "6163--6166",
booktitle = "2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2014",
address = "United States",
}