To solve the problem of calibrating the radius of a ball indenter in a hardness tester, a laser confocal radius measurement and calibration method for the ball indenter is proposed without separating the ball from the body of the indenter. The laser confocal radius measurement and calibration method uses the maximum of the confocal axial intensity curve to precisely identify the cat's eye and confocal position of the test ball indenter. The distance between these two positions is then measured to achieve high-precision radius measurement. The theoretical analyses and experimental results indicate that the radius measurement uncertainty of the ball indenter with a diameter of 1.5875 mm is within 0.12 μm.