A new reflection approach for field distribution measurement of ultrasonic transducers was investigated. Instead of a point-like reflection target (rigid sphere) or thin wires (line-like targets), a line response function of experimental knife-edge distribution combined with the inverse Abel transforms was used to estimate the lateral beam distributions of ultrasonic transducers. The measurement steps were as follows: 1 A knife-edge was scanned perpendicularly to acoustic beam axis of the transducer using an ultrasonic C-scan system to obtain its ultrasonic image line response function, 2 the transverse beam distribution was solved by the inverse Abel transforms, and 3 experiments were performed to obtain the lateral beam profiles of two transducers, with and without focus, and the results were compared with those from a hydrophone. The results showed that this method was effective for ultrasonic field measurement and could be as a substitute for hydrophone in most cases.

源语言英语
页(从-至)211-215
页数5
期刊Journal of Beijing Institute of Technology (English Edition)
20
2
出版状态已出版 - 6月 2011