Experimental study on transient high temperature measurement in explosive field

Yun Ming Zhang, Qing Ming Liu*, Lei Li, Jian Ping Wang

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

摘要

In order to meet the demands of transient high temperature measurement in explosive field, an infrared colorimetric temperature measurement system was established based on the theory of light radiation. The experiments with a blackbody furnace were carried out to calibrate the working parameters of the Si/Ge colorimetric pyrometer including wavelength, bandwidth, and photoelectric conversion coefficient, measurement range. And the upper limit of the colorimetric pyrometer reached 3500 ℃. The response time of the colorimetric pyrometer was not more than 10 μs, which was determined by calibration experiments using spark discharge. The temperature formulas were given for the standard and transfinite pyrometers. And the colorimetric pyrometer was applied to measure the real temperatures in TNT and FAE explosive fields. The outstanding features of the colorimetric pyrometer in the engineering applications indicate that the colorimetric pyrometer can be used to diagnose explosion processes and evaluate heat radiation damage effects.

源语言英语
页(从-至)513-520
页数8
期刊Baozha Yu Chongji/Expolosion and Shock Waves
34
5
DOI
出版状态已出版 - 25 9月 2014