摘要
The technical problems concerning the determination of the laser beam parameters by using area array charge coupled device (CCD) detector were studied. The beam parameters of Hermite-Gaussian modes generated from laser diode (LD) pumped CW Nd:YAG laser with 1064 nm and a lamp pumped Q-switched Nd:YAG laser with 10 ns pulse width were experimentally measured. Comparing the measured beam quality factor M2 with theoretical one, the measuring deviation of M2 was less than 5%. The measuring error of the divergence was also less than 5%. The influences of various factors on the measured beam parameters were discussed.
源语言 | 英语 |
---|---|
页(从-至) | 993-996 |
页数 | 4 |
期刊 | Zhongguo Jiguang/Chinese Journal of Lasers |
卷 | 32 |
期 | 7 |
出版状态 | 已出版 - 7月 2005 |