Endoscopic optical coherence tomography with a focus-adjustable probe

Wenchao Liao, Tianyuan Chen, Chengming Wang, Wenxin Zhang, Zhangkai Peng, Xiao Zhang, Shengnan Ai, Deyong Fu, Tieying Zhou, Ping Xue*

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

20 引用 (Scopus)
Plum Print visual indicator of research metrics
  • Citations
    • Citation Indexes: 19
  • Captures
    • Readers: 29
see details

摘要

We present a focus-adjustable endoscopic probe for optical coherence tomography (OCT), which is able to acquire images with different focal planes and overcome depth-of-focus limitations by image fusing. The use of a two-way shape-memory-alloy spring enables the probe to adjust working distance over 1.5 mm, providing a large scanning range with high resolution and no sensitivity loss. Equipped with a homemade hollow-core ultrasonic motor, the probe is capable of performing an unobstructed 360 deg field-of-view distal scanning. Both the axial resolution and the best lateral resolution are ∼4 μm, with a sensitivity of 100.3 dB. Spectral-domain OCT imaging of phantom and biological tissues with the probe is also demonstrated.

源语言英语
页(从-至)4040-4043
页数4
期刊Optics Letters
42
20
DOI
出版状态已出版 - 15 10月 2017
已对外发布

指纹

探究 'Endoscopic optical coherence tomography with a focus-adjustable probe' 的科研主题。它们共同构成独一无二的指纹。

引用此

Liao, W., Chen, T., Wang, C., Zhang, W., Peng, Z., Zhang, X., Ai, S., Fu, D., Zhou, T., & Xue, P. (2017). Endoscopic optical coherence tomography with a focus-adjustable probe. Optics Letters, 42(20), 4040-4043. https://doi.org/10.1364/OL.42.004040