Dual-Polarity Ion Trap Mass Spectrometry: Dynamic Monitoring and Controlling Gas-phase Ion–Ion Reactions

Muyi He, You Jiang, Dan Guo, Xingchuang Xiong, Xiang Fang, Wei Xu*

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

8 引用 (Scopus)

摘要

A dual-polarity linear ion trap (LIT) mass spectrometer was developed in this study, and the method for simultaneously controlling and detecting cations and anions was proposed and realized in the LIT. With the application of an additional dipolar DC field on the ejection electrodes of an LIT, dual-polarity mass spectra could be obtained, which include both the mass-to-charge (m/z) ratio and charge polarity information of an ion. Compared with conventional method, the ion ejection and detection efficiency could also be improved by about one-fold. Furthermore, ion–ion reactions within the LIT could be dynamically controlled and monitored by manipulating the distributions of ions with opposite charge polarities. This method was then used to control and study the reaction kinetics of ion–ion reactions, including electron transfer dissociation (ETD) and charge inversion reactions. A dual-polarity collision-induced dissociation (CID) experiment was proposed and performed to enhance the sequence coverage of a peptide ion. Ion trajectory simulations were also carried out for concept validation and system optimization. [Figure not available: see fulltext.].

源语言英语
页(从-至)1262-1270
页数9
期刊Journal of the American Society for Mass Spectrometry
28
7
DOI
出版状态已出版 - 1 7月 2017

指纹

探究 'Dual-Polarity Ion Trap Mass Spectrometry: Dynamic Monitoring and Controlling Gas-phase Ion–Ion Reactions' 的科研主题。它们共同构成独一无二的指纹。

引用此