摘要
A new algorithm for distortion correction based on pixel-movement model (PMM) in angiography system is proposed. The parameters of PMM are derived by analyzing the pixel movement induced by the distortion (calculated by a single-image distortion correction method) in a series of phantom images acquired by a certain interval of orientation angle. The model allows distortion correction of angiographic images acquired in any arbitrary C-arm orientation. Experimental results demonstrate that the proposed algorithm is fast with high accuracy for distortion correction of angiography images.
源语言 | 英语 |
---|---|
页(从-至) | 1473-1479 |
页数 | 7 |
期刊 | Tien Tzu Hsueh Pao/Acta Electronica Sinica |
卷 | 36 |
期 | 8 |
出版状态 | 已出版 - 8月 2008 |