摘要
A simple and sensitive reversed phase liquid chromatographic method with UV detection was developed and validated for determination of swertisin in rat plasma. Chromatographic separation was achieved on a Phenomenex Gemini C 18 column (250 × 4.6 mm, 5 μm). The mobile phase consisted of acetonitrile and water (24:76) with 1% acetic acid and was delivered at a flow rate of 0.8 mL min-1. The detection wavelength was set at 338 nm. Standard curves were linear (r ≧ 0.9976) in rat plasma over the concentration range of 0.102-10.2 μg mL-1. Within- and between-day precisions ranged from 2.6% to 6.8% RSD. The lower limit of quantification was 0.102 μg mL-1. The validated method has been successfully applied for the preliminary pharmacokinetic study of swertisin in rat.
源语言 | 英语 |
---|---|
页(从-至) | 795-797 |
页数 | 3 |
期刊 | Chromatographia |
卷 | 69 |
期 | 7-8 |
DOI | |
出版状态 | 已出版 - 4月 2009 |