Design of synchronous projection data acquisition system for cone beam CT based on manipulator

Yong You, Fei Li, Zhang Teng, Chun Guang Xu, Ming Chuan Zhang, Jing Rui Ren

科研成果: 书/报告/会议事项章节会议稿件同行评审

4 引用 (Scopus)

摘要

In order to satisfy the requirement of automatic nondestructive testing of micro defects for the casting and forging components whose shape and curvature are complex (such as aircraft engine blades), we develop a CT testing system based on six degrees of freedom industrial manipulator using computerized tomography scan of detection method. The measurement system is mainly composed of three major parts: manipulator, flat panel detector and high energy X-ray source. We carry out further analysis for the purpose of the coordinated control system and the overall design scheme based on the analysis of the basic composition and working mechanism of each unit introduced in this paper. To meet the data synchronization acquisition in every angle for the testing work piece, we design the bottom program, hardware circuit and host program to realize coordination and orderly operation of the whole detection system based on the analysis above.

源语言英语
主期刊名Proceedings of 2016 IEEE Far East NDT New Technology and Application Forum, FENDT 2016
编辑Chunguang Xu
出版商Institute of Electrical and Electronics Engineers Inc.
190-194
页数5
ISBN(电子版)9781509026630
DOI
出版状态已出版 - 25 7月 2017
活动2016 IEEE Far East NDT New Technology and Application Forum, FENDT 2016 - Nanchang, 中国
期限: 22 6月 201624 6月 2016

出版系列

姓名Proceedings of 2016 IEEE Far East NDT New Technology and Application Forum, FENDT 2016

会议

会议2016 IEEE Far East NDT New Technology and Application Forum, FENDT 2016
国家/地区中国
Nanchang
时期22/06/1624/06/16

指纹

探究 'Design of synchronous projection data acquisition system for cone beam CT based on manipulator' 的科研主题。它们共同构成独一无二的指纹。

引用此