Comparative analysis of SWIFT and Grid-SWIFT waveforms for improved ion excitation and isolation resolution

Al Fakih Ali Mohammed, Chaohong Feng*, Ting Jiang*, Wei Xu

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

摘要

Miniaturized mass spectrometers, with their increasing portability and performance enhancements, play a crucial role in selectively analyzing target compounds within complex mixtures, particularly in situ environments. As part of our ongoing efforts to enhance the analytical capabilities of the custom-built “brick” mass spectrometer, this study compared and optimized methods for ion isolation using Stored Waveform Inverse Fourier Transform (SWIFT) and Grid-SWIFT waveforms. The efficacy of the SWIFT and Grid-SWIFT waveforms in selective ion isolation was investigated, and through optimization of both waveforms, improved ion isolation efficiency, and peak resolution were demonstrated by adjusting waveform duration and amplitude. The results conclusively showed that Grid-SWIFT outperforms traditional SWIFT waveforms in enhancing the isolation of target ions. Additionally, a novel simulation-based approach was developed to simulate the behavior of ion trajectories under the SWIFT and Grid-SWIFT waveforms within LIT to provide a theoretical basis for better performance of the Grid-SWIFT waveforms. Results show that longer waveform durations help reduce energy leakage, providing higher isolation efficiency. Grid-SWIFT waveforms exhibited significantly less energy leakage compared to SWIFT waveforms. Consequently, using Grid-SWIFT as isolation waveforms improved the mass spectra' intensity and resolution.

源语言英语
文章编号127117
期刊Talanta
283
DOI
出版状态已出版 - 1 2月 2025

指纹

探究 'Comparative analysis of SWIFT and Grid-SWIFT waveforms for improved ion excitation and isolation resolution' 的科研主题。它们共同构成独一无二的指纹。

引用此