Bilateral fitting subtracting confocal microscopy

Weiqian Zhao, Zhong Sheng, Lirong Qiu*, Yun Wang, Rongjun Shao

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

11 引用 (Scopus)

摘要

This paper proposes a bilateral fitting subtracting confocal microscopy (BFSCM) based on the optical arrangement of conventional confocal microscopy (CM). BFSCM first uses the data in both sides of a confocal axial response curve, which are very sensitive to the axial position of the sample, for respective linear fitting to obtain two fitting straight lines, and then obtains a difference confocal line by subtraction of the two fitting lines. Finally, it calculates the zero position of the difference confocal line to precisely capture the focus position of the confocal system, and thereby achieving a high-precision measurement of the 3D structure of the sample. The theoretical analyses and experiments indicate that BFSCM can improve the axial resolution, and has anti-interference capability and focusing ability with bipolar absolute zero point tracking, while it does not change the structure and lateral resolution of CM. BFSCM provides a novel method for the improvement of CM axial resolution.

源语言英语
页(从-至)10269-10275
页数7
期刊Applied Optics
55
36
DOI
出版状态已出版 - 20 12月 2016

指纹

探究 'Bilateral fitting subtracting confocal microscopy' 的科研主题。它们共同构成独一无二的指纹。

引用此