A novel system for stereotactic surgery: Preliminary evaluation of targeting accuracy

Nan Xiao, Ping Luo, Shuxiang Guo, Ping Guo

科研成果: 书/报告/会议事项章节会议稿件同行评审

摘要

A new stereotactic system for deep brain stimulation has been previously proposed and its method for creating the stereotactic frame has been validated. In this study, the system was evaluated for its targeting accuracy based on a plastic skull model. A micro stereotactic frame is created based on the CT volume image acquired from the skull model. During the experiment, the frame is attached to skull-implanted screws that extend through the skull. The frame provides guidance for parallel cannulas as they are advanced through a burr hole to the target. During the experiment, a micro camera was placed in the skull model to record the whole procedure of the needle touching the target. The results showed that our stereotactic frame could fulfill the requirements of the DBS surgery.

源语言英语
主期刊名2015 IEEE International Conference on Mechatronics and Automation, ICMA 2015
出版商Institute of Electrical and Electronics Engineers Inc.
733-737
页数5
ISBN(电子版)9781479970964
DOI
出版状态已出版 - 2 9月 2015
活动12th IEEE International Conference on Mechatronics and Automation, ICMA 2015 - Beijing, 中国
期限: 2 8月 20155 8月 2015

出版系列

姓名2015 IEEE International Conference on Mechatronics and Automation, ICMA 2015

会议

会议12th IEEE International Conference on Mechatronics and Automation, ICMA 2015
国家/地区中国