A method for in vivo treatment verification of IMRT and VMAT based on electronic portal imaging device

Jun Zhang*, Xiuqing Li, Miaomiao Lu, Qilin Zhang, Xile Zhang, Ruijie Yang, Maria F. Chan, Junhai Wen*

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

5 引用 (Scopus)

摘要

Background: Intensity-modulated radiation therapy (IMRT) and volume-modulated arc therapy (VMAT) are rather complex treatment techniques and require patient-specific quality assurance procedures. Electronic portal imaging devices (EPID) are increasingly used in the verification of radiation therapy (RT). This work aims to develop a novel model to predict the EPID transmission image (TI) with fluence maps from the RT plan. The predicted TI is compared with the measured TI for in vivo treatment verification. Methods: The fluence map was extracted from the RT plan and corrections of penumbra, response, global field output, attenuation, and scatter were applied before the TI was calculated. The parameters used in the model were calculated separately for central axis and off-axis points using a series of EPID measurement data. Our model was evaluated using a CIRS thorax phantom and 20 clinical plans (10 IMRT and 10 VMAT) optimized for head and neck, breast, and rectum treatments. Results: Comparisons of the predicted and measured images were carried out using a global gamma analysis of 3%/2 mm (10% threshold) to validate the accuracy of the model. The gamma pass rates for IMRT and VMAT were greater than 97.2% and 94.5% at 3%/2 mm, respectively. Conclusion: We have developed an accurate and straightforward EPID-based quality assurance model that can potentially be used for in vivo treatment verification of the IMRT and VMAT delivery.

源语言英语
文章编号232
期刊Radiation Oncology
16
1
DOI
出版状态已出版 - 12月 2021

指纹

探究 'A method for in vivo treatment verification of IMRT and VMAT based on electronic portal imaging device' 的科研主题。它们共同构成独一无二的指纹。

引用此