A dual-source miniature mass spectrometer with improved sensitivity

Xinyu Si, Lili Hu, Wei Xu, Hang Li, Cuiping Li*

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

9 引用 (Scopus)

摘要

Previously, we have reported the development of a miniature mass spectrometer with a continuous atmospheric pressure interface, whose sensitivity was sacrificed due to the limited gas flow rate. In this study, a circular lens was designed and integrated on the ion transfer pathway between the two pumping stages, and a ten-fold sensitivity improvement could be achieved when coupling with external ionization sources. Besides the capability of working with ionization sources in the atmosphere, there is also an in-vacuum plasma ionization source, so that both volatile and non-volatile samples could be analyzed. A 10 ppbv detection sensitivity was achieved for dimethyl methyl phosphonate (DMMP) in air using the plasma ionization source. Furthermore, the miniature mass spectrometer was also demonstrated for the direct analyses of progesterone in drinking water and atrazine in soil samples using paper spray and syringe spray, respectively.

源语言英语
页(从-至)15-19
页数5
期刊International Journal of Mass Spectrometry
423
DOI
出版状态已出版 - 12月 2017