A calibration method for high-range pressure sensor used for underwater explosion

Ren Rong Long*, Qing Ming Zhang, Bi Wu, Zheng Ping Huang, Xia Chen

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

摘要

A high range hydraulic impulse calibrator was designed to calibrate the sensitivity of pressure sensor in a compact device at GPa high pressure. The compressed gas drive technique was implanted. The impact velocity of hammer was controlled by the strength of dowel which controls the pressure of driving gas. A GPa magnitude impulsive pressure with duration of several ms could be induced by the new calibrator. The calibrator was used for calibration test, and the linear regression method was used in the data processing. The results show that the experimental data are of close agreement, and are identical with the factory sensitivity coefficient of sensor. The device was proved to have good practicability and reliability.

源语言英语
页(从-至)46-49
页数4
期刊Binggong Xuebao/Acta Armamentarii
33
SUPPL.1
出版状态已出版 - 5月 2012