A 0.6V, 8.4uW AFE circuit for biomedical signal recording

Chengying Chen*, Liming Chen, Xinghua Wang, Feng Zhang

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

8 引用 (Scopus)
Plum Print visual indicator of research metrics
  • Citations
    • Citation Indexes: 8
  • Captures
    • Readers: 8
see details

摘要

For the detection of biomedical signal in wearable health devices, in this paper a fully-integrated, low-power, low-noise analog front-end (AFE) circuit is proposed. The AFE circuit includes a fully-differential amplifier with high-pass characteristics and a 14bit/500 Hz Sigma-Delta modulator based on cascode inverter. The fully-differential amplifier increases the output swing and improves the signal to noise ratio (SNR). Inverter-based Sigma-Delta modulator using bootstrapped switch completes high accuracy signal sampling in low supply voltage and finally outputs digital code. Two subthreshold bandgaps provide stable bias current and reference voltage for amplifier and Sigma-Delta modulator respectively. The circuit is fabricated with SMIC 0.13 μm CMOS mixed-signal technology occupying area of 1.68 mm2. The measurement results show that when the power supply voltage is 0.6 V, the input referred voltage noise of amplifier is 3.976μVrms and noise efficient factor (NEF) is 3.658. The maximum signal to noise and distortion ratio (SNDR) of AFE is 66.7 dB with only 8.4μw power consumption. Compared with the previous references, it has the maximum SNDR among them.

源语言英语
页(从-至)105-112
页数8
期刊Microelectronics Journal
75
DOI
出版状态已出版 - 5月 2018

指纹

探究 'A 0.6V, 8.4uW AFE circuit for biomedical signal recording' 的科研主题。它们共同构成独一无二的指纹。

引用此

Chen, C., Chen, L., Wang, X., & Zhang, F. (2018). A 0.6V, 8.4uW AFE circuit for biomedical signal recording. Microelectronics Journal, 75, 105-112. https://doi.org/10.1016/j.mejo.2018.03.009