摘要
According to the definition of contrast and contrast sensitivity function, a method of producing test targets for human eye contrast sensitivity measurement is proposed with integrating spheres and special light modulators. Then an automated measurement system is built and human eye contrast sensitivity function can be measured by psychophysical experiment with cross ladder method. Its spatial frequency covers 0.8, 1.5, 2.0, 3.0, 4.5, 6.0, 12.0 and 18.0 cycles per degree. The indication error of background luminance of test targets satisfies ±5%, and contrast repeatability is 0.001 and reproducibility is 0.004, which meet requirements of human eye contrast sensitivity measurement.
投稿的翻译标题 | Measurement Method for Contrast Sensitivity Function of the Human Eye |
---|---|
源语言 | 繁体中文 |
页(从-至) | 20-23 |
页数 | 4 |
期刊 | Jiliang Xuebao/Acta Metrologica Sinica |
卷 | 39 |
期 | 1 |
DOI | |
出版状态 | 已出版 - 22 1月 2018 |
关键词
- Contrast
- Contrast sensitivity function
- Human eye vision
- Integrating sphere
- Metrology
- Psychophysics