TY - JOUR
T1 - Study on covalent coupling process and flow characteristics of antibody on the surface of immunoassay microfluidic chip
AU - Zhong, Hao
AU - Li, Yong
AU - Liu, Guodong
AU - Xu, Tao
AU - Suo, Yiping
AU - Wang, Zhiqiang
N1 - Publisher Copyright:
© 2021 Taylor & Francis Group, LLC.
PY - 2022
Y1 - 2022
N2 - The immune response system of immunoassay microfluidic chips is a dynamic reaction process that continuously sends reactants to the surface of a solid carrier. Signal acquisition results from the heterogeneous immune reactions and reactant transport. Antibody immobilization is the most important part of heterogeneous immune reactions, and reactant transport is reflected in the form of fluid velocity. Here, we reported several surface modification processes on polystyrene substrates that are employed to study the relationship between the antibody immobilization and flow behavior in heterogeneous immune response processes. The antibody was immobilized using covalent grafting. Based on the mechanism of sandwich enzyme linked immunosorbent assay, a fluorescence quantitative detection method was used to evaluate the immune response process. The effects of different surface modification processes on immune response and flow behavior were studied. We identified an optimal flow velocity in the dynamic immune response system in the microfluidic chip. The immune response signal was the strongest when the average flow velocity was approximately 0.2 mm/s in the procalcitonin detection system. Compared with the amino and aldehyde group substrates, the epoxy group substrate has the highest antibody immobilization efficiency; compared with the surface modified by small molecular groups, the introduction of Poly-L-Lysine can increase the amount of antibody immobilization.
AB - The immune response system of immunoassay microfluidic chips is a dynamic reaction process that continuously sends reactants to the surface of a solid carrier. Signal acquisition results from the heterogeneous immune reactions and reactant transport. Antibody immobilization is the most important part of heterogeneous immune reactions, and reactant transport is reflected in the form of fluid velocity. Here, we reported several surface modification processes on polystyrene substrates that are employed to study the relationship between the antibody immobilization and flow behavior in heterogeneous immune response processes. The antibody was immobilized using covalent grafting. Based on the mechanism of sandwich enzyme linked immunosorbent assay, a fluorescence quantitative detection method was used to evaluate the immune response process. The effects of different surface modification processes on immune response and flow behavior were studied. We identified an optimal flow velocity in the dynamic immune response system in the microfluidic chip. The immune response signal was the strongest when the average flow velocity was approximately 0.2 mm/s in the procalcitonin detection system. Compared with the amino and aldehyde group substrates, the epoxy group substrate has the highest antibody immobilization efficiency; compared with the surface modified by small molecular groups, the introduction of Poly-L-Lysine can increase the amount of antibody immobilization.
KW - Flow velocity
KW - fluorescence detection
KW - microfluidic chip
KW - sandwich immune reaction
KW - surface modification
UR - http://www.scopus.com/inward/record.url?scp=85112264955&partnerID=8YFLogxK
U2 - 10.1080/10826068.2021.1958344
DO - 10.1080/10826068.2021.1958344
M3 - Article
C2 - 34383625
AN - SCOPUS:85112264955
SN - 1082-6068
VL - 52
SP - 424
EP - 432
JO - Preparative Biochemistry and Biotechnology
JF - Preparative Biochemistry and Biotechnology
IS - 4
ER -