A new method of SPR signal amplification based on dynamic field enhancement at the sensor surface to detect C-reactive protein (CRP) is proposed. The gold nanoparticles-detection antibody (AuNPs-dAb) conjugates, as the dynamical agent, were applied to enhance the biosensor sensitivity. The mechanism of SPR sensitivity enhancement is due to the strong evanescent field coupling between the localized SPR (LSPR) of AuNPs and SPR of gold-silver alloy film which proved by COMSOL Multiphysics and MATLAB simulation. The biosensor sensitivity got with sandwich immunoassay (cAb/CRP/AuNPs-dAb) is 5.56-fold of the structure without AuNPs (cAb/CRP/dAb). Moreover, concentration of AuAg-SPR biosensor in CRP detection was lower for an order of magnitude than that of the conventional gold film cladding SPR (Au-SPR) biosensor, according to its direct reaction with the immobilized cAb. The dynamic field enhancement based AuAg-SPR biosensors have high sensitivity in detecting trace biomarker of CRP, and the improvement can also be applied to the detection of other biomarkers in low concentration sensitively.
Original language | English |
---|---|
Article number | 9328342 |
Pages (from-to) | 9523-9529 |
Number of pages | 7 |
Journal | IEEE Sensors Journal |
Volume | 21 |
Issue number | 7 |
DOIs | |
Publication status | Published - 1 Apr 2021 |
Externally published | Yes |
Keywords
- Surface plasmon resonance
- dynamic field enhancement
- gold nanoparticles
- gold-silver alloy film
- localized surface plasmon resonance
- sensitivity