New calibration technology of relative transverse sensitivity of high shock tri-axis accelerometer

Ran Lin, Zhenhai Zhang*, Kejie Li, Xu He, Liang Zhang

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

2 Citations (Scopus)

Abstract

This paper examines the relative transverse sensitivity calibration technology of high shock tri-axis accelerometers. The latest established national standard normalizes the relative transverse sensitivity parameter of accelerometers. While this is an important index, it applies only to single dimensional accelerometers. To extend the index to calibrations of high shock tri-axis accelerometers, a method was adopted to collect data and calibrations based on an air gun, using linear line fitting to replace the calibration curve and computing relative transverse sensitivity with division of lateral chip output and main axis chip, which makes it more convenient for users to evaluate performance. This will perfect the testing and calibrating method of relative transverse sensitivity for the tri-axis accelerometer. Compared with others, the adopted calibration device and method not only showed high peak value and wide pulse duration, but also presented the complete output performance of three axes.

Original languageEnglish
Pages (from-to)922-928
Number of pages7
JournalZhendong Ceshi Yu Zhenduan/Journal of Vibration, Measurement and Diagnosis
Volume36
Issue number5
DOIs
Publication statusPublished - 1 Oct 2016

Keywords

  • Air gun
  • Calibration
  • High shock
  • Relative transverse sensitivity
  • Tri-axis accelerometer

Fingerprint

Dive into the research topics of 'New calibration technology of relative transverse sensitivity of high shock tri-axis accelerometer'. Together they form a unique fingerprint.

Cite this