A gradient shimming method is presented to promote the homogeneity, and the full width at half maximum (FWHM) of the resonance line is used to measure the effect of gradient shimming. Results of experiments, showed the FWHM as a basic criterion for gradient shimming, which has worked efficiently in the 0.4 T magnetic resonance imaging (MRI) scanner.

Original languageEnglish
Pages (from-to)279-282
Number of pages4
JournalBeijing Ligong Daxue Xuebao/Transaction of Beijing Institute of Technology
Volume29
Issue number3
Publication statusPublished - Mar 2009