ADRC-based control method for the vascular intervention master–slave surgical robotic system

Wei Zhou, Shuxiang Guo*, Jin Guo*, Fanxu Meng, Zhengyang Chen

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

24 Citations (Scopus)

Abstract

In vascular interventional surgery, surgeons operate guidewires and catheters to diagnose and treat patients with the assistance of the digital subtraction angiography (DSA). Therefore, the surgeon will be exposed to X-rays for extended periods. To protect the surgeon, the development of a robot-assisted surgical system is of great significance. The displacement tracking accuracy is the most important issue to be considered in the development of the system. In this study, the active disturbance rejection control (ADRC) method is applied to guarantee displacement tracking accuracy. First, the core contents of the proportional–integral–derivative (PID) and ADRC methods are analyzed. Second, comparative evaluation experiments for incremental PID and ADRC methods are presented. The results show that the ADRC method has better performance of than that of the incremental PID method. Finally, the calibration experiments for the ADRC control method are implemented using the master–slave robotic system. These experiments demonstrate that the maximum tracking error is 0.87 mm using the ADRC method, effectively guaranteeing surgical safety.

Original languageEnglish
Article number1439
JournalMicromachines
Volume12
Issue number12
DOIs
Publication statusPublished - Dec 2021

Keywords

  • ADRC-based control method
  • Active disturbance rejection control (ADRC)
  • Master–slave robotic system
  • Surgical safety
  • Vascular interventional surgery

Fingerprint

Dive into the research topics of 'ADRC-based control method for the vascular intervention master–slave surgical robotic system'. Together they form a unique fingerprint.

Cite this