Abstract
We proposed and experimentally demonstrated a short-delayed self-heterodyne method with 15.5m delay to get a large-frequency-range laser frequency-noise spectrum over 10Hz to 50 MHz, and an averaging approach to extract the intrinsic frequency noise of a frequency-swept laser. With these two techniques, dynamic frequency-noise spectrum of a frequency-swept DFB laser when free running and servo-controlled are both measured. This measurement method permits accurate and insightful investigation of laser stability.
Original language | English |
---|---|
Pages (from-to) | 29245-29257 |
Number of pages | 13 |
Journal | Optics Express |
Volume | 23 |
Issue number | 22 |
DOIs | |
Publication status | Published - 2 Nov 2015 |
Externally published | Yes |