Development of an electronic portal imaging device dosimetry method

Jun Zhang, Ziting Fan, Xile Zhang, Ruijie Yang, Junhai Wen*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

3 Citations (Scopus)

Abstract

Support arm backscatter and off-axis effects of an electronic portal imaging device (EPID) are challenging for radiotherapy quality assurance. Aiming at the issue, we proposed a simple yet effective method with correction matrices to rectify backscatter and off-axis responses for EPID images. First, we measured the square fields with ionization chamber array (ICA) and EPID simultaneously. Second, we calculated the dose-to-pixel value ratio and used it as the correction matrix of the corresponding field. Third, the correction value of the large field was replaced with that of the same point in the small field to generate a correction matrix suitable for different EPID images. Finally, we rectified the EPID image with the correction matrix, and then the processed EPID images were converted into the absolute dose. The calculated dose was compared with the measured dose via ICA. The gamma pass rates of 3%/3 mm and 2%/2 mm (5% threshold) were 99.6% ± 0.94% and 95.48% ± 1.03%, and the average gamma values were 0.28 ± 0.04 and 0.42 ± 0.05, respectively. Experimental results verified our method accurately corrected EPID images and converted pixel values into absolute dose values such that EPID was an efficient radiotherapy dosimetry tool.

Original languageEnglish
Article number1654
JournalDiagnostics
Volume11
Issue number9
DOIs
Publication statusPublished - Sept 2021

Keywords

  • Backscatter
  • EPID
  • Off-axis
  • Portal dosimetry
  • Quality assurance
  • Radiotherapy

Fingerprint

Dive into the research topics of 'Development of an electronic portal imaging device dosimetry method'. Together they form a unique fingerprint.

Cite this