Scanner-Independent MyoMapNet for Accelerated Cardiac MRI T1 Mapping Across Vendors and Field Strengths

  • Amine Amyar
  • , Ahmed S. Fahmy
  • , Rui Guo
  • , Kei Nakata
  • , Eiryu Sai
  • , Jennifer Rodriguez
  • , Julia Cirillo
  • , Karishma Pareek
  • , Jiwon Kim
  • , Robert M. Judd
  • , Frederick L. Ruberg
  • , Jonathan W. Weinsaft
  • , Reza Nezafat*
  • *此作品的通讯作者
  • Harvard University
  • Boston University
  • Cornell University
  • Duke University

科研成果: 期刊稿件文章同行评审

摘要

Background: In cardiac T1 mapping, a series of T1-weighted (T1w) images are collected and numerically fitted to a two or three-parameter model of the signal recovery to estimate voxel-wise T1 values. To reduce the scan time, one can collect fewer T1w images, albeit at the cost of precision or/and accuracy. Recently, the feasibility of using a neural network instead of conventional two- or three-parameter fit modeling has been demonstrated. However, prior studies used data from a single vendor and field strength; therefore, the generalizability of the models has not been established. Purpose: To develop and evaluate an accelerated cardiac T1 mapping approach based on MyoMapNet, a convolution neural network T1 estimator that can be used across different vendors and field strengths by incorporating the relevant scanner information as additional inputs to the model. Study Type: Retrospective, multicenter. Population: A total of 1423 patients with known or suspected cardiac disease (808 male, 57 ± 16 years), from three centers, two vendors (Siemens, Philips), and two field strengths (1.5 T, 3 T). The data were randomly split into 60% training, 20% validation, and 20% testing. Field Strength/Sequence: A 1.5 T and 3 T, Modified Look-Locker inversion recovery (MOLLI) for native and postcontrast T1. Assessment: Scanner-independent MyoMapNet (SI-MyoMapNet) was developed by altering the deep learning (DL) architecture of MyoMapNet to incorporate scanner vendor and field strength as inputs. Epicardial and endocardial contours and blood pool (by manually drawing a large region of interest in the blood pool) of the left ventricle were manually delineated by three readers, with 2, 8, and 9 years of experience, and SI-MyoMapNet myocardial and blood pool T1 values (calculated from four T1w images) were compared with conventional MOLLI T1 values (calculated from 8 to 11 T1w images). Statistical Tests: Equivalency test with 95% confidence interval (CI), linear regression slope, Pearson correlation coefficient (r), Bland–Altman analysis. Results: The proposed SI-MyoMapNet successfully created T1 maps. Native and postcontrast T1 values measured from SI-MyoMapNet were strongly correlated with MOLLI, despite using only four T1w images, at both field-strengths and vendors (all r > 0.86). For native T1, SI-MyoMapNet and MOLLI were in good agreement for myocardial and blood T1 values in institution 1 (myocardium: 5 msec, 95% CI [3, 8]; blood: −10 msec, 95%CI [−16, −4]), in institution 2 (myocardium: 6 msec, 95% CI [0, 11]; blood: 0 msec, [−18, 17]), and in institution 3 (myocardium: 7 msec, 95% CI [−8, 22]; blood: 8 msec, [−14, 30]). Similar results were observed for postcontrast T1. Data Conclusion: Inclusion of field strength and vendor as additional inputs to the DL architecture allows generalizability of MyoMapNet across different vendors or field strength. Evidence Level: 2. Technical Efficacy: Stage 2.

源语言英语
页(从-至)179-189
页数11
期刊Journal of Magnetic Resonance Imaging
59
1
DOI
出版状态已出版 - 1月 2024
已对外发布

学术指纹

探究 'Scanner-Independent MyoMapNet for Accelerated Cardiac MRI T1 Mapping Across Vendors and Field Strengths' 的科研主题。它们共同构成独一无二的学术指纹。

引用此