Probe alignment and design issues of microelectromechanical system based optical coherence tomography endoscopic imaging

Can Duan*, Jingjing Sun, Sean Samuelson, Huikai Xie

*此作品的通讯作者

科研成果: 期刊稿件文章同行评审

17 引用 (Scopus)

摘要

Endoscopic optical coherence tomography (OCT) imaging has been demonstrated using microelectromechanical system (MEMS) technology by several research groups. The focus of this work is to study how the OCT imaging performance is affected by the radius of curvature of MEMS mirrors as well as the optical alignment accuracy inside small imaging probes. The goal of this study is to provide guidance for assembly tolerance and design optimization of OCT endoscopic probes. Gaussian beam propagation is used for theoretical analysis which is confirmed by optical simulation and verified experimentally with a time-domain OCT system as well. It has been found that the OCT imaging is very sensitive to the distance from the fiber end to the gradient-index (GRIN) lens, which needs to be controlled within 0.1 mm to achieve working distance (WD) longer than 3.5 mm and lateral resolution around 25 μm. The impact on image quality of the MEMS mirror is negligible if the radius of curvature of the mirror surface is greater than 200 mm. In addition, we studied the astigmatismintroduced by cylindrical plastic tubing; the maximum astigmatism ratio is 1.1 when the WD is around 2.5 mm.

源语言英语
页(从-至)6589-6598
页数10
期刊Applied Optics
52
26
DOI
出版状态已出版 - 10 9月 2013
已对外发布

指纹

探究 'Probe alignment and design issues of microelectromechanical system based optical coherence tomography endoscopic imaging' 的科研主题。它们共同构成独一无二的指纹。