跳到主要导航 跳到搜索 跳到主要内容

Physical symmetry and virtual plane-based reduction reference: A preliminary study for robot-assisted pelvic fracture reduction

  • Lifeng Wang
  • , Tianmiao Wang
  • , Changsheng Li
  • , Peifu Tang
  • , Ying Xu
  • , Lihai Zhang
  • , Na Guo
  • , Yanpeng Zhao
  • , Lu Zhao
  • , Lei Hu*
  • *此作品的通讯作者
  • Beihang University
  • General Hospital of People's Liberation Army

科研成果: 期刊稿件 › 文章 › 同行评审

摘要

Traditional pelvis fracture reduction suffers from some disadvantages. Robot-assisted pelvis fracture reduction offers some promise in solving these problems. However, the reduction reference to guide robot motion is a key issue that must be resolved. In this paper, we propose a physical symmetry and virtual plane-based reduction reference and adopt the method of registration to calculate the virtual plane for the reference, which were verified via experiments. The results of the position symmetry experiments of the original pelvis and virtual plane-based position symmetry experiments were similar; both showed that the symmetry errors of the pelvis were less than 4mm and 2.5?. The results indicated that the proposed method could be used as a reference for robot-assisted pelvis fracture reduction.

源语言英语
期刊论文编号1640014
期刊Journal of Mechanics in Medicine and Biology
卷16
期8
DOI
出版状态已出版 - 1 12月 2016
已对外发布是