跳到主要导航 跳到搜索 跳到主要内容

Ophthalmic fundus camera design based on freeform surface for reducing refractive error sensitivity

  • Wenchao Zhang
  • , Weilin Chen
  • , Jun Chang*
  • , Yi Huang
  • , Xuehui Zhao
  • , Xuyang Li
  • *此作品的通讯作者
  • Beijing Institute of Technology
  • CAS - Xi'an Institute of Optics and Precision Mechanics

科研成果: 期刊稿件文章同行评审

摘要

The eye has a remarkably complex structure and biomechanics. An imbalance between the forces acting on several muscles and the tissue properties may alter or even preclude vision. With the widespread use of electronics, refractive errors have become a common problem, making clear fundus imaging difficult. To overcome this challenge, freeform surfaces have emerged as a potential solution, enabling compact, low-cost, and user-friendly systems that are insensitive to refractive errors. We propose a fundus camera based on a freeform surface that can image the fundus without the influence of refractive errors. A special front lens was designed to image the pupil on a freeform lens that can realize phase modulation. The system performs well providing a field of view of 40 and a pupil diameter of 3 mm for refractive errors ranging from -5D to +5D. The volume of the system was less than 35 mm × 35 mm × 135 mm.

源语言英语
文章编号107714
期刊Optics and Lasers in Engineering
169
DOI
出版状态已出版 - 10月 2023
已对外发布

指纹

探究 'Ophthalmic fundus camera design based on freeform surface for reducing refractive error sensitivity' 的科研主题。它们共同构成独一无二的指纹。

引用此