跳到主要导航 跳到搜索 跳到主要内容

Novel optical system for scanning system of biochip

  • Qian Wang*
  • , Lin Li
  • , Lian Sheng An
  • , Yi Fan Huang
  • *此作品的通讯作者
  • Beijing Institute of Technology

科研成果: 期刊稿件 › 文章 › 同行评审

摘要

The conventional fluorescent confocal scanning system has high resolution, big aperture, and short working distance. The working distances usually are about 1mm, and some are even shorter, which have limit in some kinds of biochip, such as cell-chip, tissue-chip and so on. Because these biochips need wide distance to real-time add medicament or reagent in it. The wider the distance is the more convenient to add the medicament or reagent. A novel optical system for scanning system of biochip is presented, which has long working distance(15 mm), big numeric aperture (0.67) and high resolution(350 nm). An optical structure is selected firstly. Through the way of CAD, a well corrected objective lens has obtained finally.

源语言英语
页(从-至)269-271
页数3
期刊Guangxue Jishu/Optical Technique
卷31
期2
出版状态已出版 - 3月 2005
已对外发布是

学术指纹

探究 'Novel optical system for scanning system of biochip' 的科研主题。它们共同构成独一无二的学术指纹。

引用此