Non-contact detection of oxygen saturation based on visible light imaging device using ambient light

  • Lingqin Kong
  • , Yuejin Zhao
  • , Liquan Dong
  • , Yiyun Jian
  • , Xiaoli Jin
  • , Bing Li
  • , Yun Feng
  • , Ming Liu
  • , Xiaohua Liu
  • , Hong Wu
  • Beijing Institute of Technology

科研成果: 期刊稿件 › 文章 › 同行评审

摘要

A method that remotely measures blood oxygen saturation through two cameras under regular lighting is proposed and experimentally demonstrated. Two narrow-band filters with their visible wavelength of 660nm and 520nm are mounted to two cameras respectively, which are then used to capture two photoplethysmographic (PPG) from the subject simultaneously. The data gathered from this system, including both blood oxygen saturation and heart rate, is compared to the output of a traditional figure blood volume pulse (BVP) senor that was employed on the subject at the same time. Result of the comparison showed that the data from the new, non-contact system is consistent and comparable with the BVP senor. Compared to other camera-based measuring method, which requires additional close-up lighting, this new method is achievable under regular lighting condition, therefore more stable and easier to implement. This is the first demonstration of an accurate video-based method for non-contact oxygen saturation measurements by using ambient light with their respective visible wavelength of 660nm and 520nm which is free from interference of the light in other bands.

源语言英语
页(从-至)17464-17471
页数8
期刊Optics Express
卷21
期15
DOI
出版状态已出版 - 29 7月 2013

学术指纹

探究 'Non-contact detection of oxygen saturation based on visible light imaging device using ambient light' 的科研主题。它们共同构成独一无二的学术指纹。

引用此