跳到主要导航 跳到搜索 跳到主要内容

Lens refractive index measurement based on fiber point-diffraction longitudinal interferometry

  • Beijing Institute of Technology

科研成果: 期刊稿件 › 文章 › 同行评审

摘要

A non-immersive lens refractive index measurement method based on fiber point-diffraction longitudinal interferometry is presented. The lens imaging process is simplified to the single refraction of the back surface if the object point is located at the vertex of the front surface. The lens refractive index is derived through measuring its thickness, radius of curvature of the back surface, the distance between the object point and the image point. Experiments indicate its accuracy is better than 2.2 × 10-4. Since the front surface is excluded in the imaging process, even an aspherical lens could be accurately measured by this method.

源语言英语
页(从-至)22389-22399
页数11
期刊Optics Express
卷21
期19
DOI
出版状态已出版 - 23 9月 2013
已对外发布是

学术指纹

探究 'Lens refractive index measurement based on fiber point-diffraction longitudinal interferometry' 的科研主题。它们共同构成独一无二的学术指纹。

引用此